Friday, 21 February 2014

ELISA (Enzyme-Linked ImmunoSorbant Assay): Troubleshoout Problem solving tips









1. Positive results in negative control

Contamination of reagents/samples
May be contamination of reagents or samples, or cross contamination from splashing between wells. Use fresh reagents and pipette carefully.
Sandwich ELISA – detection antibody is detecting coating antibody
Check the correct coating antibody and detection antibodies are being used and that they will not detect each other.
Insufficient washing of plates
Ensure well areas are washed adequately by filling the wells with wash buffer. Ensure all residual antibody solutions are removed before washing.
Too much antibody used leading to non-specific binding
Check the recommended amount of antibody suggested. Try using less antibody.

2. High background across entire plate

Conjugate to strong or left on too long
Check dilution of conjugate, use it at the recommended dilution. Stop the reaction using stop buffer as soon as the plate has developed enough for absorbance readings.
Substrate solution or stop solution is not fresh
Use fresh substrate solution. Stop solution should be clear (if it has gone yellow, this is a sign of contamination and it should be replaced).
Reaction not stopped
Color will keep developing if the substrate reaction is not stopped.
Plate left too long before reading on the plate reader
Color will keep developing (though at a slower rate if stop solution has been added).
Contaminants from laboratory glassware
Ensure reagents are fresh and prepared in clean glassware. Sterilize glassware beforehand if possible.
Substrate incubation carried out in the light
Substrate incubation should be carried out in the dark.
Incubation temperature too high
Antibodies will have optimum binding activity at the correct temperature. Ensure the incubations are carried out at the correct temperature and that incubators are set at the correct temperature and working. Incubation temperature may require some optimization.
Non-specific binding of antibody
Ensure a block step is included and a suitable blocking buffer is being used. We recommend using 5-10 % serum from the same species of the secondary antibody, or bovine serum.
Ensure wells are pre-processed to prevent non specific attachment. Use an affinity purified antibody, preferably pre-absorbed.
Also check suggestions listed under ‘Positive results in negative control’

3. Low absorbance values

Target protein not expressed in sample used/ Low level of target protein expression in sample used
Check the expression profile of the target protein to ensure it will be expressed in your samples. If there is low level of target protein expression, increase the amount of sample used, or you may need to change to a more sensitive assay. Ensure you are using a positive control within the detection range of the assay.
Insufficient antibody
Check the recommended amount of antibody is being used. The concentration of antibody may require increasing for optimization of results.
Substrate solutions not fresh or combined incorrectly
Prepare the substrate solutions immediately before use. Ensure the stock solutions are in date and have been stored correctly, and are being used at the correct concentration. Ensure the reagents are used as directed at the correct concentration.
Reagents not fresh or not at the correct pH
Ensure reagents have been prepared correctly and are in date.
Incubation time not long enough
Ensure you are incubating the antibody for the recommended amount of time, if an incubation time is suggested. The incubation time may require increasing for optimization of results.
Incubation temperature too low
Antibodies will have optimum binding activity at the correct temperature. Ensure the incubations are carried out at the correct temperature and that incubators are set at the correct temperature and working. Incubation temperature may require some optimization. Ensure all reagents are at room temperature before proceeding.
Stop solution not added
Addition of stop solution increases the intensity of color reaction and stabilizes the final color reaction.

4. High absorbance values for samples and/or positive control (absorbance does not go down as the sample is diluted down the plate)

The concentration of samples or positive control is too high and out of range for the sensitivity of the assay. Re-assess the assay you are using OR reduce the concentration of samples and control by dilution before adding to the plate. Take the dilution into account when calculating the resulting concentrations.

5. Inconsistent absorbances across the plate

Plates stacked during incubations
Stacking of plates does not allow even distribution of temperature across the wells of the plates. Avoid stacking.
Pipetting inconsistent
Ensure pipettes are working correctly and are calibrated. Ensure pipette tips are pushed on far enough to create a good seal. Take particular care when diluting down the plate and watch to make sure the pipette tips are all picking up and releasing the correct amount of liquid. This will greatly affect consistency of results between duplicates.
Antibody dilutions/reagents not well mixed
To ensure a consistent concentration across all wells, ensure all reagents and samples are mixed before pipetting onto the plate.
Wells allowed to dry out
Ensure lids are left on the plates at all times when incubating. Place a humidifying water tray (bottled clean/sterile water) in the bottom of the incubator.
Inadequate washing
This will lead to some wells not being washed as well as others, leaving different amounts of unbound antibody behind which will give inconsistent results.
Bottom of the plate is dirty affecting absorbance readings
Clean the bottom of the plate carefully before re-reading the plate.

6. Color developing slowly

Plates are not at the correct temperature
Ensure plates are at room temperature and that the reagents are at room temperature before use.
Conjugate too weak
Prepare the substrate solutions immediately before use. Ensure the stock solutions are in date and have been stored correctly, and are being used at the correct concentration. Ensure the reagents are used as directed, at the correct concentration.
Contamination of solutions
Presence of contaminants, such as sodium azide and peroxidase can affect the substrate reaction. Avoid using reagents containing these preservatives.




1 comment:

  1. Wow, What a Excellent post. I really found this to much informatics. It is what i was searching for.I would like to suggest you that please keep sharing such type of info.Thanks viza canada din romana

    ReplyDelete